  • Site > Home > Medical Monitor > UT-2000 Patient Monitor
  • Product images
  • UT-2000 Patient Monitor

  • Detailed description
  • UT-2000 Patient Monitor


    Product Function

    1.20 Type Arrhythmic Analysis, Multi-Lead ECG Waveforms Display in Phase, Real time S_T segment analysis, pacemaker detection Drug calculation and titration table

    2. Efficient resistance to interference of defibrillator and electrosurgical cautery

    3. SPO2 can testing for 0.1% Weak

    4. RA-LL impedance Respiration

    5. Trend Coexist Display

    6. OxyCRG DyNamic View Display

    8. Networking capacity and nurse_calling system

    9. Options of printing, IBP, VGA and EtCO2, touch screen.

    10. UP to 4 hours working capacity of built_in rechargeable battery.

    11, 12.1"high resolution color TFT LCD display.

    12, Large volume of tabular and graphic trends information storage and easy to recall.

    13, Anti-ESU, anti- defibrillator

    14, 6 Standard parameters:ECG,RESP,NIBP,SPO2,1_TEMP,PR/HR.

    15, Capture dynamic waveforms.

     

    Product Specification

    ECG

    Lead Mode: 5 Leads (       R, L, F, N, C or RA, LA, LL, RL,V)

    Lead selection: I, II, III, avR, avL, avF, V,

    Waveform: 2 ch

    Lead mode: 3 Leads ( R, L, F or RA, LA, LL)

    Lead selection: I, II, III,

    Waveform: 1 ch

    Gain: ´2.5mm/mV, ´5.0mm/mV, ´10mm/mV, ´20mm/mV,  auto

    HR and Alarm             

    Range   

    Adult: 15 ~ 300 bpm

    Neo/Ped: 15 ~ 350 bpm

    Accuracy       : ± 1% or ± 1bpm,which great

    Resolution: 1 bpm

    Sensitivity: > 200 (uV P-P)

    Differential Input Impedance: > 5 MΩ

    CMRR

    Monitor: > 105 dB

    Operation: > 105 dB

    Diagnosis: > 85 dB

    Electrode offset potential: ±300mV

    Leakage Current: < 10 uA

    Baseline Recovery: < 3 S  After Defi.

    ECG Signal Range: ±8 m V ( Vp-p )      

    Bandwidth  

    Surgery: 1 ~ 15 Hz

    Monitor: 0.5 ~ 35 Hz

    Diagnostic: 0.05 ~ 100 Hz

    Calibration Signal: 1 (mV p-p),  Accuracy:±5% 

    ST Segment Monitoring Range: Measure and Alarm -2.0 ~ +2.0 mV

    ARR Detecting    

    Type: ASYSTOLE, VFIB/VTAC, COUPLET, BIGEMINY, TRIGEMINY, R ON T, VT>2, PVC, TACHY, BRADY, MISSED BEATS, PNP, PNC

    Alarm: Available

    Review: Available

    RESPARATION

    Method: Impedance between R-F(RA-LL)

    Differential Input Impedance: >2.5 MΩ

    Measuring Impedance Range:  0.3~5.0Ω

    Base line Impedance Range: 0 – 2.5 

    Bandwidth: 0.3 ~ 2.5 Hz

    Resp. Rate

    Measuring and Alarm Range   

    Adult: 0 ~ 120 rpm

    Neo/Ped: 0 ~ 150 rpm

    Resolution: 1 rpm

    Accuracy       : ±2 rpm

    Apean Alarm: 10 ~ 40 S

    NIBP

    Method: Oscillometric

    Mode: Manual, Auto, STAT

    Measuring Interval in AUTO Mode: 1, 2, 3, 4, 5, 10, 15, 30, 60, 90, 120, 180, 240,480 (Min)

    Measuring Period in STAT Mode: 5 Min

    Pulse Rate Range: 40 ~ 240 bpm

    Alarm Type: SYS, DIA, MEAN

    Measuring and alarm range                   

      Adult Mode

    SYS: 40 ~ 270 mmHg

    DIA: 10 ~ 215 mmHg

    MEAN: 20 ~ 235 mmHg

      Pediatric Mode

    SYS: 40 ~ 200 mmHg

    DIA: 10 ~ 150 mmHg

        MEAN:20 ~ 165 mmHg

    Neonatal Mode

    SYS: 40 ~ 135 mmHg

    DIA:10 ~ 100 mmHg

    MEAN:20 ~ 110 mmHg

    Resolution Pressure:1mmHg

    Accuracy       Pressure  Maximum Mean error: ±5mmHg

    Maximum Standard deviation: ±8mmHg

    Overpressure Protection

    Adult Mode: 297±3 mmHg

    Pediatric Mode: 240±3 mmHg

    Neonatal Mode: 147±3 mmHg

    SpO2

    Measuring Range: 0 ~ 100 %

    Alarm Range: 0 ~ 100 %

    Resolution: 1 %

    Accuracy       : 70% ~ 100% ±2 %

         0% ~ 69% unspecified

    Actualization interval: about 1 Sec.

    Alarm Delay: 10 Sec.

    Pulse Rate

    Measuring and Alarm Range: 20~300bpm

    Resolution: 1bpm

    Accuracy       : ±2bpm

    TEMPERATURE

    Channel: 1

    Measuring and Alarm Range: 0 ~ 50 °C

    Resolution: 0.1°C

    Accuracy       : ±0.1°C

    Actualization interval: about 1 Sec.

    Average Time Constant:<10 Sec.

    IBPOptional

    Label: ART, PA, CVP, RAP, LAP, ICP, P1, P2

    Measuring and alarm range

    ART: 0 ~ 300 mmHg

           PA: -6 ~ 120 mmHg

           CVP/RAP/LAP/ICP: -10 ~ 40 mmHg

           P1/P2: -10 ~ 300 mmHg                                

    Press Sensor

    Sensitivity: 5 uV/V/mmHg

    Impedance: 300-3000Ω

    Resolution: 1 mmHg

    Accuracy       : ±2% or ±1mmHg, which great

    Actualization interval: about 1 Sec.

    Packing List

     

    Standard

    UT2000 Patient Monitor main unit (without printer)

    1 pc

    5-lead ECG cable                                   

    1pc

    Adult finger SpO2 sensor                             

    1 pc

    Adult NIBP cuff                                    

    1 pc

    NIBP extension tube                                 

    1 pc

    TEMP Skin Probe                                    

    1 pc

    Power Wire                                        

    1 pc

    The User manual

    1 pc

     